The 90 series is designed for rugged field applications in wet environments. The case, keypad and connectors are rated to IP65.
To comply with GLP guidelines, the date, time and results of the last calibration are stored in memory, along with the unit's serial number. This data can be displayed or sent directly to the RS232 port. Warning of failed calibration is provided. All readings stored in memory are stamped with the date and time.
A standard feature of the 90 series is the Notepad function. Up to 9340 readings can be recorded for later review. These can be downloaded to a printer or computer via the RS232 port. For added functionality, the user is able to enter up to two additional items to be tagged on to the reading. Many operators use this function as a pond or site identifier, depth or readings from other instruments.
Automatic datalogging provides the capability to automatically record up to 9340 sets of readings at user-set intervals. The unit remains dormant between readings and only wakes up whenever a reading is due. Battery life is considerably improved with this facility.
All recorded data can be downloaded directly to a Printer or Computer with the standard RS232 port (cable sold separately). With the WinTPS communication software, many instrument functions can be controlled from the computer. This makes the units ideal for remote monitoring.
Product Group | No |
---|---|
Technical Information | pH Range : 0.00 to 14.00 pH Resolution : 0.01 pH Accuracy : ±0.01pH Impedance : >3 x 1012 Ohms Asymmetry : -1.00 to +1.00 pH Range : 85 to 105%
Temperature Compensation : Automatic, -5 to 100OC Calibration : 2 Point automatic pH Buffers : Auto recognition of pH4.01, pH6.86, pH7.00, pH9.18 and pH10.01. Other buffers can be keyed in at calibration.
Millivolts Range : 0 to +/-1999 mV Resolution : 1 mV Accuracy : +/-1 mV
Temperature Range : -10.0 to 110.0 OC Resolution : 0.1 OC Accuracy : +/-0.2 OC Calibration : Automatic, 1 or 2 point
Good Laboratory Practices Date, time and value of last calibration for each parameter are stored. This information can be recalled or sent to the RS232 port at any time.
Warning of failed calibration is provided.
All readings stored in memory are stamped with the date and time.
Datalogging Memory Capacity 9340 readings (7230 with extra data input active).
Extra Data Input 2 extra items of data can be input when using the Notepad function. Each item can be a maximum of 4 characters, including a decimal point. Extra Data items are labelled A and B.
Automatic Logging The 90-P can be programmed to... - Automatically log from 1 to 288 readings per day. - Automatically log readings at a preset interval for a preset duration.
RS232 Serial Port Output 8 Bits, No Parity, 1 Stop Bit, XON/OFF Protocol. 300, 9600 & 19200 baud available.
Display 2 line x 40 Character alphanumeric LCD shows all readings simultaneously.
Keypad 5 Function keys, Menu key, 14 data entry keys, OFF, ON.
Dimensions 230 x 140 x 100 mm
Mass Meter only : Approx 1.5 kg Full Kit : Approx 5.0 kg
Environment Temperature : 0 to 45°C Humidity : 0 to 90% R.H.
Power 7.2V Rechargeable NiCad Battery. Any 12VDC supply will recharge the battery. 123145/1 |
Pack size | KIT |
pH
Range : 0.00 to 14.00 pH
Resolution : 0.01 pH
Accuracy : ±0.01pH
Impedance : >3 x 1012 Ohms
Asymmetry : -1.00 to +1.00 pH
Range : 85 to 105%
Temperature Compensation : Automatic, -5 to 100OC Calibration : 2 Point automatic pH Buffers : Auto recognition of pH4.01, pH6.86, pH7.00, pH9.18 and pH10.01. Other buffers can be keyed in at calibration.
Millivolts
Range : 0 to +/-1999 mV
Resolution : 1 mV
Accuracy : +/-1 mV
Temperature
Range : -10.0 to 110.0 OC
Resolution : 0.1 OC
Accuracy : +/-0.2 OC
Calibration : Automatic, 1 or 2 point
Good Laboratory Practices
Date, time and value of last calibration for each parameter are stored. This information can be recalled or sent to the RS232 port at any time.
Warning of failed calibration is provided.
All readings stored in memory are stamped with the date and time.
Datalogging
Memory Capacity
9340 readings (7230 with extra data input active).
Extra Data Input
2 extra items of data can be input when using the Notepad function. Each item can be a maximum of 4 characters, including a decimal point. Extra Data items are labelled A and B.
Automatic Logging
The 90-P can be programmed to...
- Automatically log from 1 to 288 readings per day.
- Automatically log readings at a preset interval for a preset duration.
RS232 Serial Port Output
8 Bits, No Parity, 1 Stop Bit, XON/OFF Protocol.
300, 9600 & 19200 baud available.
Display
2 line x 40 Character alphanumeric LCD shows all readings simultaneously.
Keypad
5 Function keys, Menu key, 14 data entry keys, OFF, ON.
Dimensions
230 x 140 x 100 mm
Mass
Meter only : Approx 1.5 kg
Full Kit : Approx 5.0 kg
Environment
Temperature : 0 to 45°C
Humidity : 0 to 90% R.H.
Power
7.2V Rechargeable NiCad Battery.
Any 12VDC supply will recharge the battery.
123145/1