Product Description
Additional Information
Technical Information
The Aqua-CP/A is a waterproof, all-in-one meter for Conductivity, TDS, pH and Temperature. It features multiple Conductivity and TDS ranges, making it suitable for a wide variety of lab and field applications. The ranges are automatically selected to provide the best available resolution.
Genuine one-touch calibration puts an end to time-consuming calibration procedures.
The Aqua-CP/A comes with all the accessories needed for immediate use. An optional hard carry case is also available.
Rugged and Reliable
TPS has used the latest microprocessor electronics to totally eliminate all internal and external adjustments. The result is a dramatic increase in reliability. The unit is ideal for field or factory use.
24 Character Alphanumeric Display
The two-line, 24 character display shows the Conductivity/TDS, pH and Temperature readings simultaneously.
Alternatively, a Large Digit mode is available. This shows one reading at a time at almost double the size. To simplify use of the Aqua-CP/A, plain text prompts and error messages are provided. There are no codes to interpret like some other brands.
Temperature Compensation
Automatic Temperature Compensation for all parameters is provided via a temperature sensor in the Conductivity/TDS sensor.
One-touch Calibration
Calibration of the Aqua-CP/A could not be easier. Just place the relevant electrode into any of the pre-selected standards and press the Calibrate key. It's that simple !
The Aqua-CP/A displays probe condition after calibration.
All calibration information is stored in memory when the Aqua-CP/A is switched off, even if the battery is removed.
Low Battery Drain
The Aqua-CP/A will provide in excess of 70 hours of continuous use with an alkaline 9 volt battery.
A handy battery-saver function switches the Aqua-CP/A off after 5 minutes if no keys have been pressed. This function can also be disabled for continuous operation.
Australian Quality
The Aqua Series is proudly designed and manufactured in Australia. Our aim is to provide you with the best quality, service and value for money. The TPS Quality System is certified in accordance with the AS/NZS ISO 9001 standard. Since 1968, TPS has built a reputation for excellent backup service that is second to none.
NOTE: 1m KIT INCLUDES
pH Sensor, 1m Surface only......121207
k=1.0 Cond-ATC-Temp 1m, plastic.122201
pH7.00 Buffer, 200mL............121387
pH4.00 Buffer, 200mL............121381
2.76 mS/cm Cond Standard, 200mL.122306
9V Alkaline Battery ............130026
Manual..........................130050
Genuine one-touch calibration puts an end to time-consuming calibration procedures.
The Aqua-CP/A comes with all the accessories needed for immediate use. An optional hard carry case is also available.
Rugged and Reliable
TPS has used the latest microprocessor electronics to totally eliminate all internal and external adjustments. The result is a dramatic increase in reliability. The unit is ideal for field or factory use.
24 Character Alphanumeric Display
The two-line, 24 character display shows the Conductivity/TDS, pH and Temperature readings simultaneously.
Alternatively, a Large Digit mode is available. This shows one reading at a time at almost double the size. To simplify use of the Aqua-CP/A, plain text prompts and error messages are provided. There are no codes to interpret like some other brands.
Temperature Compensation
Automatic Temperature Compensation for all parameters is provided via a temperature sensor in the Conductivity/TDS sensor.
One-touch Calibration
Calibration of the Aqua-CP/A could not be easier. Just place the relevant electrode into any of the pre-selected standards and press the Calibrate key. It's that simple !
The Aqua-CP/A displays probe condition after calibration.
All calibration information is stored in memory when the Aqua-CP/A is switched off, even if the battery is removed.
Low Battery Drain
The Aqua-CP/A will provide in excess of 70 hours of continuous use with an alkaline 9 volt battery.
A handy battery-saver function switches the Aqua-CP/A off after 5 minutes if no keys have been pressed. This function can also be disabled for continuous operation.
Australian Quality
The Aqua Series is proudly designed and manufactured in Australia. Our aim is to provide you with the best quality, service and value for money. The TPS Quality System is certified in accordance with the AS/NZS ISO 9001 standard. Since 1968, TPS has built a reputation for excellent backup service that is second to none.
NOTE: 1m KIT INCLUDES
pH Sensor, 1m Surface only......121207
k=1.0 Cond-ATC-Temp 1m, plastic.122201
pH7.00 Buffer, 200mL............121387
pH4.00 Buffer, 200mL............121381
2.76 mS/cm Cond Standard, 200mL.122306
9V Alkaline Battery ............130026
Manual..........................130050
Product Group | No |
---|---|
Technical Information | Conductivity with k=0.1 sensor Ranges : 0 to 2.000 uS/cm 0 to 20.00 uS/cm 0 to 200.0 uS/cm 0 to 2000 uS/cm Resolution : 0.001 uS/cm 0.01 uS/cm 0.1 uS/cm 1 uS/cm 0.01 mS/cm Accuracy : +/-0.2% of full scale at 25 OC with k=1.0 sensor Ranges : 0 to 20.00 uS/cm 0 to 200.0 uS/cm 0 to 2000 uS/cm 0 to 20.00 mS/cm Resolution : 0.01 uS/cm 0.1 uS/cm 1 uS/cm 0.01 mS/cm 0.1 mS/cm Accuracy : +/-0.2% of full scale at 25 OC with k=10 sensor Ranges : 0 to 200.0 uS/cm 0 to 2000 uS/cm 0 to 20.00 mS/cm 0 to 200.0 mS/cm Resolution : 0.1 uS/cm 1 uS/cm 0.01 mS/cm 0.1 mS/cm 1 mS/cm Accuracy : +/-0.2% of full scale at 25 OC TDS with k=0.1 sensor Ranges : 0 to 1.000 ppM 0 to 10.00 ppM 0 to 100.0 ppM 0 to 1000 ppM Resolution : 0.001 ppM 0.01 ppM 0.1 ppM 1 ppM 0.01 ppK Accuracy : +/-0.3% of full scale at 25 OC with k=1.0 sensor Ranges : 0 to 10.00 ppM 0 to 100.0 ppM 0 to 1000 ppM 0 to 10.00 ppK Resolution : 0.01 ppM 0.1 ppM 1 ppM 0.01 ppK 0.1 ppK Accuracy : +/-0.3% of full scale at 25 OC with k=10 sensor Ranges : 0 to 100.0 ppM 0 to 1000 ppM 0 to 10.00 ppK 0 to 100.0 ppK Resolution : 0.1 ppM 1 ppM 0.01 ppK 0.1 ppK 1 ppK Accuracy : +/-0.3% of full scale at 25 OC pH Range : 0 to 14.00 pH Resolution : 0.01 pH Accuracy : +/-0.01pH Temperature Range : -10.0 to 120.0 OC (Cond sensor limit 60 OC) Resolution : 0.1 OC Accuracy : +/-0.2 OC pH Input Impedance >3 x 1012 Ohms pH Asymmetry Range -1.00 to +1.00 pH pH Slope Range 85.0 to 105.0% Temperature Compensation Conductivity Automatic, 0 to 100 OC pH Automatic or manual, 0 to 100 OC Automatic Standard Recognition Conductivity : 150 uS/cm 1413 uS/cm 2.76 mS/cm 12.88 mS/cm 58.0 mS/cm TDS : 69.5 ppM 2.00 ppK 8.00 ppK 36.0 ppK pH 4.00, 6.88, 7.00, 9.23 and 10.06 pH Enclosure Tough, ABS plastic case, waterproof to IP65. Display 24 Character alphanumeric LCD with full-text prompts and error messages. Battery Saver Function On : Automatically switches off after 5 minutes of no use. Off : Unit remains on for continuous use. Power 9V Alkaline battery for 70+ hours. Dimensions 157 x 78 x 35 mm Mass Instrument only : Approx 250g Full Kit : Approx 2.5kg Environment Temperature : 0 to 45 OC Humidity : 0 to 90% R.H. 121135/1 121135/K1/1 121135/1/K1 |
Pack size | KIT |
Conductivity
with k=0.1 sensor
Ranges : 0 to 2.000 uS/cm
0 to 20.00 uS/cm
0 to 200.0 uS/cm
0 to 2000 uS/cm
Resolution : 0.001 uS/cm
0.01 uS/cm
0.1 uS/cm
1 uS/cm
0.01 mS/cm
Accuracy : +/-0.2% of full scale
at 25 OC
with k=1.0 sensor
Ranges : 0 to 20.00 uS/cm
0 to 200.0 uS/cm
0 to 2000 uS/cm
0 to 20.00 mS/cm
Resolution : 0.01 uS/cm
0.1 uS/cm
1 uS/cm
0.01 mS/cm
0.1 mS/cm
Accuracy : +/-0.2% of full scale
at 25 OC
with k=10 sensor
Ranges : 0 to 200.0 uS/cm
0 to 2000 uS/cm
0 to 20.00 mS/cm
0 to 200.0 mS/cm
Resolution : 0.1 uS/cm
1 uS/cm
0.01 mS/cm
0.1 mS/cm
1 mS/cm
Accuracy : +/-0.2% of full scale
at 25 OC
TDS
with k=0.1 sensor
Ranges : 0 to 1.000 ppM
0 to 10.00 ppM
0 to 100.0 ppM
0 to 1000 ppM
Resolution : 0.001 ppM
0.01 ppM
0.1 ppM
1 ppM
0.01 ppK
Accuracy : +/-0.3% of full scale
at 25 OC
with k=1.0 sensor
Ranges : 0 to 10.00 ppM
0 to 100.0 ppM
0 to 1000 ppM
0 to 10.00 ppK
Resolution : 0.01 ppM
0.1 ppM
1 ppM
0.01 ppK
0.1 ppK
Accuracy : +/-0.3% of full scale
at 25 OC
with k=10 sensor
Ranges : 0 to 100.0 ppM
0 to 1000 ppM
0 to 10.00 ppK
0 to 100.0 ppK
Resolution : 0.1 ppM
1 ppM
0.01 ppK
0.1 ppK
1 ppK
Accuracy : +/-0.3% of full scale
at 25 OC
pH
Range : 0 to 14.00 pH
Resolution : 0.01 pH
Accuracy : +/-0.01pH
Temperature
Range : -10.0 to 120.0 OC
(Cond sensor limit 60 OC)
Resolution : 0.1 OC
Accuracy : +/-0.2 OC
pH Input Impedance
>3 x 1012 Ohms
pH Asymmetry Range
-1.00 to +1.00 pH
pH Slope Range
85.0 to 105.0%
Temperature Compensation
Conductivity
Automatic, 0 to 100 OC
pH
Automatic or manual, 0 to 100 OC
Automatic Standard Recognition
Conductivity : 150 uS/cm
1413 uS/cm
2.76 mS/cm
12.88 mS/cm
58.0 mS/cm
TDS : 69.5 ppM
2.00 ppK
8.00 ppK
36.0 ppK
pH
4.00, 6.88, 7.00, 9.23 and 10.06 pH
Enclosure
Tough, ABS plastic case, waterproof to IP65.
Display
24 Character alphanumeric LCD with full-text prompts and error messages.
Battery Saver Function
On : Automatically switches off
after 5 minutes of no use.
Off : Unit remains on for
continuous use.
Power
9V Alkaline battery for 70+ hours.
Dimensions
157 x 78 x 35 mm
Mass
Instrument only : Approx 250g
Full Kit : Approx 2.5kg
Environment
Temperature : 0 to 45 OC
Humidity : 0 to 90% R.H.
121135/1 121135/K1/1 121135/1/K1
with k=0.1 sensor
Ranges : 0 to 2.000 uS/cm
0 to 20.00 uS/cm
0 to 200.0 uS/cm
0 to 2000 uS/cm
Resolution : 0.001 uS/cm
0.01 uS/cm
0.1 uS/cm
1 uS/cm
0.01 mS/cm
Accuracy : +/-0.2% of full scale
at 25 OC
with k=1.0 sensor
Ranges : 0 to 20.00 uS/cm
0 to 200.0 uS/cm
0 to 2000 uS/cm
0 to 20.00 mS/cm
Resolution : 0.01 uS/cm
0.1 uS/cm
1 uS/cm
0.01 mS/cm
0.1 mS/cm
Accuracy : +/-0.2% of full scale
at 25 OC
with k=10 sensor
Ranges : 0 to 200.0 uS/cm
0 to 2000 uS/cm
0 to 20.00 mS/cm
0 to 200.0 mS/cm
Resolution : 0.1 uS/cm
1 uS/cm
0.01 mS/cm
0.1 mS/cm
1 mS/cm
Accuracy : +/-0.2% of full scale
at 25 OC
TDS
with k=0.1 sensor
Ranges : 0 to 1.000 ppM
0 to 10.00 ppM
0 to 100.0 ppM
0 to 1000 ppM
Resolution : 0.001 ppM
0.01 ppM
0.1 ppM
1 ppM
0.01 ppK
Accuracy : +/-0.3% of full scale
at 25 OC
with k=1.0 sensor
Ranges : 0 to 10.00 ppM
0 to 100.0 ppM
0 to 1000 ppM
0 to 10.00 ppK
Resolution : 0.01 ppM
0.1 ppM
1 ppM
0.01 ppK
0.1 ppK
Accuracy : +/-0.3% of full scale
at 25 OC
with k=10 sensor
Ranges : 0 to 100.0 ppM
0 to 1000 ppM
0 to 10.00 ppK
0 to 100.0 ppK
Resolution : 0.1 ppM
1 ppM
0.01 ppK
0.1 ppK
1 ppK
Accuracy : +/-0.3% of full scale
at 25 OC
pH
Range : 0 to 14.00 pH
Resolution : 0.01 pH
Accuracy : +/-0.01pH
Temperature
Range : -10.0 to 120.0 OC
(Cond sensor limit 60 OC)
Resolution : 0.1 OC
Accuracy : +/-0.2 OC
pH Input Impedance
>3 x 1012 Ohms
pH Asymmetry Range
-1.00 to +1.00 pH
pH Slope Range
85.0 to 105.0%
Temperature Compensation
Conductivity
Automatic, 0 to 100 OC
pH
Automatic or manual, 0 to 100 OC
Automatic Standard Recognition
Conductivity : 150 uS/cm
1413 uS/cm
2.76 mS/cm
12.88 mS/cm
58.0 mS/cm
TDS : 69.5 ppM
2.00 ppK
8.00 ppK
36.0 ppK
pH
4.00, 6.88, 7.00, 9.23 and 10.06 pH
Enclosure
Tough, ABS plastic case, waterproof to IP65.
Display
24 Character alphanumeric LCD with full-text prompts and error messages.
Battery Saver Function
On : Automatically switches off
after 5 minutes of no use.
Off : Unit remains on for
continuous use.
Power
9V Alkaline battery for 70+ hours.
Dimensions
157 x 78 x 35 mm
Mass
Instrument only : Approx 250g
Full Kit : Approx 2.5kg
Environment
Temperature : 0 to 45 OC
Humidity : 0 to 90% R.H.