Product Group | No |
---|---|
Technical Information | Conductivity Ranges : 4 ranges with auto range selection k=0.1 Sensor : 0 to 2.000 uS/cm 0 to 20.00 uS/cm 0 to 200.0 uS/cm 0 to 2000 uS/cm k=1 Sensor : 0 to 20.00 uS/cm 0 to 200.0 uS/cm 0 to 2000 uS/cm 0 to 20.00 mS/cm k=10 Sensor : 0 to 200.0 uS/cm 0 to 2000 uS/cm 0 to 20.00 mS/cm 0 to 200.0 mS/cm Resolution : 0.05% of range Accuracy : +/-0.5% of full scale of selected range at 25OC
TDS Ranges : 4 ranges with auto range selection. k=0.1 Sensor : 0 to 1.000 ppM 0 to 10.00 ppM 0 to 100.0 ppM 0 to 1000 ppM k=1 Sensor : 0 to 10.00 ppM 0 to 100.0 ppM 0 to 1000 ppM 0 to 10.00 ppK k=10 Sensor : 0 to 100.0 ppM 0 to 1000 ppM 0 to 10.00 ppK 0 to 100.0 ppK Resolution : 0.1% of range Accuracy : +/-0.5% of full scale of selected range at 25OC
Temperature Range : -10.0 to 120.0 OC (sensor limit 60 OC) Resolution : 0.1 OC Accuracy : +/-0.2 OC Conductivity/TDS Span Range
75% to 133% of nominal k factor
Temperature Compensation
Automatic, 0 to 60 OC.
Automatic Standards Recognition
Conductivity :
150 uS/cm
1413 uS/cm
2.76 mS/cm
12.88 mS/cm
58.0 mS/cm
TDS :
69.5 ppM
2.00 ppK
8.00 ppK
36.0 ppK
Enclosure
Tough, anti-static Luran-S plastic case, waterproof to IP67. Connectors waterproof to IP65.
Display
32 Character alphanumeric LCD with user-friendly menu system and full-text messages.
Memory
2400 readings, including date and time.
Automatic Logging
User-set for one reading every 1 to 90 seconds, 1 to 90 minutes or 1 to 24 hours.
RS232 Serial Port Output
300, 1200 and 9600 Baud.
8 Bits, No Parity, 1 Stop Bit, XON/XOFF
Clock
24 Hour Calendar Clock
Battery Saver Function
On : Automatically switches off after 5 minutes of no use.
Off : Unit remains on for continuous use.
Good Laboratory Practices
Date, time and results of last successful calibration are stored, and can be recalled when required. GLP information can be printed or downloaded to PC with the RS232 serial port interface. Warning of failed calibration is provided.
Power
6V NiMH Rechargeable battery pack for approx 75 hours operation. Battery charger for country of choice supplied. Battery can be charged from any 12V DC source.
Dimensions
187 x 110 x 51 mm
Mass
Instrument only : Approx 440g
Full Kit : Approx 2.5kg
Environment
Temperature : 0 to 45 OC
Humidity : 0 to 95% R.H. |
Pack size | EACH |
Conductivity
Ranges : 4 ranges with auto range selection
k=0.1 Sensor :
0 to 2.000 uS/cm
0 to 20.00 uS/cm
0 to 200.0 uS/cm
0 to 2000 uS/cm
k=1 Sensor :
0 to 20.00 uS/cm
0 to 200.0 uS/cm
0 to 2000 uS/cm
0 to 20.00 mS/cm
k=10 Sensor :
0 to 200.0 uS/cm
0 to 2000 uS/cm
0 to 20.00 mS/cm
0 to 200.0 mS/cm
Resolution : 0.05% of range
Accuracy : +/-0.5% of full scale of selected range at 25OC
TDS
Ranges : 4 ranges with auto range selection.
k=0.1 Sensor :
0 to 1.000 ppM
0 to 10.00 ppM
0 to 100.0 ppM
0 to 1000 ppM
k=1 Sensor :
0 to 10.00 ppM
0 to 100.0 ppM
0 to 1000 ppM
0 to 10.00 ppK
k=10 Sensor :
0 to 100.0 ppM
0 to 1000 ppM
0 to 10.00 ppK
0 to 100.0 ppK
Resolution : 0.1% of range
Accuracy : +/-0.5% of full scale of selected range at 25OC
Temperature
Range : -10.0 to 120.0 OC (sensor limit 60 OC)
Resolution : 0.1 OC
Accuracy : +/-0.2 OC