Product Categories
- Apparatus
- Chemicals
-
Equipment
- Autoclaves
- Balances
- Centrifuges
- Colony Counters
- Dispersers
- Dry Block Heaters
- Environmental Chambers
- Filtration
- Freezers
- Fume Cupboards
- Furnaces
- Grinders & Mills
- Specials
- Second Hand
All miniCHEM units are easy to install. The enclosure is waterproof to IP65, so a separate cabinet is not required. All controls can be accessed via the hinged, clear cover. The large LCD display is easy to read, even from a distance.
Product Group | No |
---|---|
Technical Information | Ranges 0 to 199.9 % Saturation OR 0 to 19.99 ppM (mg/L) (Factory-set)
Resolution 0.1 % Saturation OR 0.01 ppM (mg/L)
Accuracy +/-0.2% of full scale
Linearity +/-0.1%
Repeatability +/-0.2%
Ambient Drift <0.05% span/OC
Long Term Drift <0.2% per year
Temperature Compensation Dual automatic temperature compensation system,(sensor limit 45 OC).
Zero Range +/-10%
Span Range 70 to 130%
Sensor YSI5739 Polarographic sensor with inbuilt ATC sensor.
Enclosure Polycarbonate, waterproof to IP65
Display 12.5mm LCD
Outputs 4 to 20 mA (500 Ohms max load) 0 to 1 V DC (0 to 10 V DC optional) Contacts : One optional relay with changeover contacts. Rated to 240V, 1 Amp
Isolation Galvanic isolation of sensor input Current and voltage outputs are not isolated
Power 220/240 V AC, 50/60Hz (110/120 V AC available) 12V DC power optional.
Dimensions 172 x 160 x 71mm
Mass Instrument only : Approx 800g Full Kit : Approx 1.5kg
Instrument Operating Environment Temperature : 0 to 45°C Humidity : 0 to 95% R.H. 113140/5 |
Pack size | KIT |
Ranges
0 to 199.9 % Saturation
OR 0 to 19.99 ppM (mg/L) (Factory-set)
Resolution
0.1 % Saturation
OR 0.01 ppM (mg/L)
Accuracy
+/-0.2% of full scale
Linearity
+/-0.1%
Repeatability
+/-0.2%
Ambient Drift
<0.05% span/OC
Long Term Drift
<0.2% per year
Temperature Compensation
Dual automatic temperature
compensation system,(sensor limit 45 OC).
Zero Range
+/-10%
Span Range
70 to 130%
Sensor
YSI5739 Polarographic sensor with inbuilt ATC sensor.
Enclosure
Polycarbonate, waterproof to IP65
Display
12.5mm LCD
Outputs
4 to 20 mA (500 Ohms max load)
0 to 1 V DC (0 to 10 V DC optional)
Contacts : One optional relay with changeover contacts. Rated to 240V, 1 Amp
Isolation
Galvanic isolation of sensor input
Current and voltage outputs are not isolated
Power
220/240 V AC, 50/60Hz
(110/120 V AC available)
12V DC power optional.
Dimensions
172 x 160 x 71mm
Mass
Instrument only : Approx 800g
Full Kit : Approx 1.5kg
Instrument Operating Environment
Temperature : 0 to 45°C
Humidity : 0 to 95% R.H.
113140/5