The smartCHEM-C is no ordinary Conductivity/TDS meter. In addition to the stunning array of features listed below, it also offers Resistivity readout, and a Ratio mode which shows the ratio of two sensors.
The built in sensor replatiniser allows the operator to maintain their sensors without having to return them to the factory.
Datalogging
The smartCHEM-C comes with a 2300 reading memory. Readings can be recorded manually, or automatically logged at preset intervals.
RS232 Computer Interface
The capabilities of the smartCHEM-C have been expanded with the RS232 port, which is standard equipment (cable sold separately). All readings stored in memory can be downloaded to an RS232 printer or PC. The RS232 port also allows the smartCHEM-C to log directly to the printer or PC. WinTPS Communication software for Windows 3.1 or 95, 98 & ME is available.
Automatic Stability Function (ASF)
ASF adds an extra level of versatility to the smartCHEM-C. When ASF is activated, the smartCHEM-C monitors all parameters that are currently in use. When ALL parameters become stable, the readings are frozen on the display. ASF is automatically activated with extra fine settings during calibration, to ensure the most accurate possible results for the user.
ASF can also be used during Manual and Automatic Datalogging. When used during Automatic Datalogging, the smartCHEM-C will log until a stable end-point is reached for all parameters, then stop logging.
Good Laboratory Practices
To comply with GLP guidelines, the date, time and results of the last calibration are stored in memory, along with the instrument’s serial number. This data can be displayed or sent to the RS232 port. All readings stored in memory are stamped with the date and time. Warning of failed calibration is provided. A calibration alarm can be programmed to whichever parameters the user chooses. The alarm can be set for Daily, Weekly, 2 Weekly or 4 Weekly. The calibration alarm can also be switched off.
PIN Code Access
The Access Code is system is provided for those users who need to ensure the integrity of recorded data, calibration settings and so forth. The Access Code function will prevent anyone who does not know the code from entering the menu system. Keys that are available during normal measurement, (Store, Print, Autolog and F1 to F5 ) are still available without the Access Code. This means that others are still able to carry out day-to-day work.
Easy to use
The smartCHEM-C has a 240 x 64 dot graphic display. It features a user-friendly menu system, making the manual virtually unnecessary. Full-text help and error messages are provided. All readings and messages are provided in large, bold characters.
Automatic Calibration
Where necessary, the user is able to define the standards to be used for calibration.
Reliability and Service
Product Group | No |
---|---|
Technical Information | Temperature
Range : -10.0 to 120.0 OC
Resolution : 0.1 OC
Accuracy : +/-0.2 OC
Temperature Compensation
Automatic, 0 to 100 OC.
User-selectable ATC coefficient,
0 to -6.0%/OC
Calibration
- Automatic calibration via Menu system.
- Any Conductivity, TDS, or Resistivity standard can be entered for calibration.
Conductivity Sensor Span Range
±25% of nominal k factor
Temperature Sensor Offset Range
-10.0 OC to +10.0 OC
Memory
2300 readings including date and time.
Automatic Datalogging
User-set for one reading every 2 to 90 seconds, minutes or hours.
RS232 Serial Port Output
1200, 9600, 19200 and 38400 Baud.
8 Bits, No Parity, 1 Stop Bit, XON/XOFF Protocol
Recorder Output (optional)
Conductivity...: 0 to 2000 mV for all ranges
TDS............: 0 to 1000 mV for all ranges
Resistivity....: 0 to 2000 mV
Ratio..........: 0 to 1500 mV
Manual range selection or range hold function should be used to maintain logical output.
Output impedance approx 1000 Ohms. Resolution approx 2 mV.
Clock
24 Hour Calendar clock.
Displays date, month, year, hours, minutes, seconds.
Good Laboratory Practices
Date, time and value of last successful calibration for each parameter are stored. This information can be recalled or sent to the RS232 port at any time.
Calibration Alarm
Calibration Alarm can be activated for parameter of user's choice. Calibration Alarm can be set to Daily, Weekly, 2 Weekly or 4 Weekly.
Display
240 x 64 dot graphic LCD with large, bold digits, user-friendly menu system and full-text help and error messages.
Enclosure
Tough, high impact ABS. Splash resistant one-piece lid.
Dimensions
240 x 180 x 105 mm
Mass
Instrument only : Approx 0.8 kg
Full Kit : Approx 3.0 kg
Environment
Temperature : 5 to 45°C
Humidity : 0 to 90% R.H.
Power
12V DC via AC/DC adaptor.
AC/DC adaptor to suit specified country is supplied.
122105 |
Pack size | EACH |