Call: (07) 38461311
0 item(s)

WP81 3m Bundle

Be the first to review this product

TPS

Part No. 648947

CAS No.

Details :


  • Rugged and Waterproof

  • User-friendly

  • Calendar Clock

  • Good Laboratory Practices

  • 2400 Reading Notepad

  • Automatic Datalogging

  • RS232 Serial Port Interface

  • Multiple Power Options

  • Australian Made

Quantity:
Product Description Additional Information Technical Information
TDS Range: 0 to 100ppK
Resolution :
0.1% of range
Accuracy :
0.5% of full scale of selected range
pH
Range : 0 to 14.00 pH
Resolution : 0.01 pH
Accuracy : +/-0.01pH
mV
Range : 0 to +/-500.0 and 0 to +/-1500 mV (auto-ranging)
Resolution : 0.15 & 1 mV
Accuracy : +/-0.3 & +/-1 mV

Temperature
Range : -10.0 to 120.0 OC
Resolution : 0.1 OC
Accuracy : +/-0.2 0OC
Conductivity/TDS Span Range
75% to 133% of nominal k factor
pH Input Impedance
>3 x 1012 Ohms pH Asymmetry Range
-1.00 to +1.00 pH
pH Slope Range
85.0 to 105.0%
Temperature Compensation
Automatic, 0 to 100 OC.
Automatic Buffer Recognition
4.00, 6.88, 7.00, 9.23 and 10.06 pH
Enclosure
Tough, anti-static Luran-S plastic case, waterproof to IP67. Connectors waterproof to IP65. The instrument will float if dropped into water.
Display
32 Character alphanumeric LCD with user-friendly menu system and full-text messages.
Memory
2400 readings, including date and time.
Automatic Logging
User-set for one reading every 1 to 90 seconds, 1 to 90 minutes or 1 to 24 hours.
RS232 Serial Port Output
300, 1200 and 9600 Baud.
8 Bits, No Parity, 1 Stop Bit, XON/XOFF
Clock
24 Hour Calendar Clock
Battery Saver Function
On : Automatically switches off after 5 minutes of no use.
Off : Unit remains on for continuous use.
Good Laboratory Practices
Date, time and results of last successful calibration are stored, and can be recalled when required. GLP information can be printed or downloaded to PC when the optional RS232 Interface is fitted. Warning of failed calibration is provided.
Power
6V NiMH Rechargeable battery pack for approx 75 hours operation. Battery charger for country of choice supplied. Battery can be charged from any 12V DC source.
Dimensions
187 x 110 x 51 mm
Mass
Instrument only : Approx 440g
Full Kit : Approx 2.5kg
Environment Temperature : 0 to 45°C
Humidity : 0 to 95% R.H.
121132/3
© 2018 Science Essentials
follow us
[profiler]
Memory usage: real: 25165824, emalloc: 22641704